Emerson Maximum Working, Pressure12, Static Pressure Limits, Rosemount 3051N, Reference Manual

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 57
Image 57
Maximum Working

Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

Figure 5-1. Typical Smart Transmitter Response Time

Transmitter 4–20 mA Output vs. Time

Pressure Released

 

 

Td

Tc

Td = Dead Time

20mA

 

 

Tc = Time Constant

 

 

 

 

Response Time =

 

 

Td+Tc

9.89mA

 

63.2% of Total

 

 

Step Change

4mA

 

Time

 

 

Maximum Working

Rosemount 3051ND

Pressure(1)(2)

Static pressure limit

Static Pressure Limits

Rosemount 3051NG and Rosemount 3051NA

Upper range limit

Operates within specifications between static line pressures stated below

Rosemount 3051ND only

Range Code

Static Pressure Limits

 

 

00.5 psia to 750 psig (3,4 kPa abs to 5,2 MPa)

10.5 psia to 2000 psig (3,4 kPa abs to 13,8 MPa)

2 – 5(1)

0.5 psia to 3626 psig (3,4 kPa abs to 25 MPa)

Overpressure Limits

Burst Pressure

Transmitters withstand following overpressure without damage:

Rosemount 3051ND/NG

Range Code

Overpressure Limits

 

 

0750 psig (5,2 MPa)

12000 psig (13,8 MPa)

2 – 5(1)

3626 psig (25 MPa)

Rosemount 3051NA

 

 

Range Code

Overpressure Limits

 

 

060 psia (414 kPa)

1120 psia (827 kPa)

2300 psia (2 070 kPa)

31600 psia (11 030 kPa)

4(1)

6000 psia (41 370 kPa)

Minimum burst pressure is 10,000 psig (69 MPa)

(1)EPR process O-ring (Code B) is limited to 2000 psi maximum working pressure.

(2)See Table 5-1for specified Seismic Maximum Working Pressure.

5-7

Page 57
Image 57
Emerson manual Maximum Working, Pressure12, Static Pressure Limits, Overpressure Limits Burst Pressure, Rosemount 3051N