Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

Section 5 Specifications and Reference Data

NUCLEAR SPECIFICATIONS

Seismic

Nuclear Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1 Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page 5-3 Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page 5-5 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-8 Materials of Construction . . . . . . . . . . . . . . . . . . . . . . . . . page 5-8 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-10 Configuration Information . . . . . . . . . . . . . . . . . . . . . . . . . page 5-11

Qualified for nuclear use per IEEE Std 344-1987 and IEEE Std 323-1983 (mild environment) as documented in Rosemount Report D2001019

Table 5-1. Seismic Specifications Summary

 

 

Range

During Seismic

 

Specified Seismic Maximum

Structural

Model

 

Code

Accuracy(1)

Post Seismic Accuracy

Working Pressure

Integrity

3051ND

 

0

not specified

not specified

750 psi (5,2 MPa)

Maintained

 

 

 

 

 

 

 

 

 

1(2)

0.75% of URL

0.25% of span

2000 psi (13,8 MPa)

throughout

 

 

 

(adjustable damping

 

 

specified seismic

 

Differential

 

≥ 1.6 s)

 

 

disturbance

 

 

 

 

 

 

 

 

0.8 s)

 

 

 

 

 

2(2)

0.75% of URL

 

3000 psi (20,7 MPa)

 

 

 

 

(adjustable damping ≥

 

(glass-filled TFE o-ring)

 

 

 

 

 

 

2000 psi (13,8 MPa)

 

 

 

3

0.75% of URL

 

 

 

 

 

 

 

(EP o-ring)

 

 

 

4

0.25% of URL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

 

 

 

 

 

 

3051NG

 

2(2)

0.75% of URL

 

Upper Range Limit

 

 

Gauge

 

(adjustable damping

 

 

 

 

 

≥ 0.8 s)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

0.75% of URL

 

 

 

 

 

 

 

 

 

 

 

 

4

0.25% of URL

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

 

 

 

 

 

 

 

 

 

3051NA

 

0

not specified

not specified

 

 

 

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

2

0.25% of URL

0.25% of span

 

 

 

 

 

 

 

 

 

 

Absolute

3

 

 

 

 

 

 

 

 

 

 

 

4

 

 

3000 psia

 

 

 

 

 

 

 

 

 

 

 

(glass-filled TFE o-ring)

 

 

 

 

 

 

2000 psi (13,8 MPa)

 

 

 

 

 

 

(EP o-ring)

 

 

 

 

 

 

 

 

(1)User-adjustable damping set at ≥ 0.4 s unless otherwise noted.

(2)Mounting bracket (Option Code B2, BS, or PM) required for specified “During Seismic Accuracy” performance.

www.rosemountnuclear.com

Page 51
Image 51
Emerson 3051N manual Nuclear Specifications, Seismic