Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

Message

 

Description

 

 

 

Add item for ALL

 

Asks the user whether the hot key item being added should be

device types or only

 

added for all device types or only for the type of device that is

for this ONE device

 

connected.

type.

 

 

 

 

 

Command Not

 

The connected device does not support this function.

Implemented

 

 

 

 

 

Communication

 

The communicator and the device are not communicating

Error

 

correctly. Check all connections between the communicator

 

 

and the device and resend the information.

 

 

 

Configuration

 

The configuration stored in memory is incompatible with the

memory not

 

device to which a transfer has been requested.

compatible with

 

 

connected device

 

 

 

 

 

CPU board not

 

The electronics board is not initialized. Return to Rosemount

initialized ON

 

Nuclear Instruments, Inc. for repair.

 

 

 

CPU EEPROM write

 

Message sent to electronics board from HART signal failed.

failure ON

 

Return to Rosemount Nuclear Instruments, Inc. for repair.

 

 

 

Device Busy

 

The connected device is busy performing another task.

 

 

 

Device

 

The device failed to respond to a command. Check all

Disconnected

 

connections between the communicator and the device and

 

 

resend the command.

 

 

 

Device write

 

Device is in write-protect mode. Data can not be written.

protected

 

 

 

 

 

Device write

 

Device is in write-protect mode. Press YES to turn the HART

protected. Do you

 

communicator off and lose the unsent data.

still want to

 

 

shut off?

 

 

 

 

 

Display value of

 

Asks whether the value of the variable should be displayed

variable on hotkey

 

adjacent to its label on the hotkey menu if the item being added

menu?

 

to the hotkey menu is a variable.

 

 

 

Download data from

 

Press the SEND softkey to transfer information from the

configuration

 

communicator memory to the device.

memory to device

 

 

 

 

 

Exceed field width

 

Indicates that the field width for the current arithmetic variable

 

 

exceeds the device-specified description edit format.

 

 

 

Exceed precision

 

Indicates that the precision for the current arithmetic variable

 

 

exceeds the device-specified description edit format.

 

 

 

Ignore next 50

 

Select YES to ignore the next 50 occurrences of device status,

occurrences of

 

or select no to display every occurrence.

status?

 

 

 

 

 

Illegal character

 

An invalid character for the variable type was entered.

 

 

 

Illegal date

 

The day portion of the date is invalid.

 

 

 

Illegal month

 

The month portion of the date is invalid.

 

 

 

Illegal year

 

The year portion of the date is invalid.

 

 

 

Incompatible CPU

 

Return to Rosemount Nuclear Instruments, Inc. for repair.

board and module

 

 

ON

 

 

 

 

 

Incomplete

 

The exponent of a scientific notation floating point variable is

exponent

 

incomplete.

 

 

 

Incomplete field

 

The value entered is not complete for the variable type.

 

 

 

A-15

Page 91