Rosemount 3051N
FIGURE 2-7. Transmitter Data
Flow with Calibration Options.
Sensor
Signal
Sensor
Input
Pressure
Reference Manual
June 2008
Transmitter Ranged 0 to 100 inH2O |
| |
Transmitter Electronics Module |
| |
Microprocessor |
|
|
Digital PV |
|
|
HART Communicator | Output Device | Analog Output |
| 20.00 mA |
|
Input Device |
|
|
Online
1 ➡ Device Setup
2 | PV | 100.00 inH2O |
3 | AO | 20.00 mA |
4 | LRV | 0.00 inH2O |
5 | URV | 100.00 inH2O |
NOTE
Value on PV line should equal the input pressure. Value on AO line should equal the output device reading.
Deciding Which Trim Procedure to Use
Sensor Trim
To decide which trim procedure to use, you must first determine whether the
1.Connect a pressure source, a HART Communicator, and a digital readout device to the transmitter.
2.Establish communication between the transmitter and the communicator.
3.Apply pressure equal to the upper range point pressure (100 inH20, for example).
4.Compare the applied pressure to the Process Variable (PV) line on the Communicator
5.Compare the Analog Output (AO) line on the communicator
You can trim the sensor using either the full trim or the zero trim function. The trim functions vary in complexity, and their use is
A zero trim is a