Emerson 3051N manual Functional Specifications, Service, Output, Power Supply, Temperature Limits

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 55
Image 55
FUNCTIONAL

Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

FUNCTIONAL

SPECIFICATIONS

Service

Liquid, gas, or vapor

Output

4–20 mA, user-selectable for linear or square root output; digital signal based

 

on HART® protocol

Power Supply

Load Limitations

 

Maximum loop resistance is determined by the voltage level of the

 

external power supply, as described by:

 

Max. Loop Resistance = 43.5 (Power Supply Voltage – 10.5) ohms

Load (ohms)

1935

 

 

 

 

 

 

 

1500

 

 

 

 

 

 

 

 

 

 

 

 

 

1000

 

 

 

 

 

Design

 

 

 

 

 

 

500

 

 

 

 

 

Region

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10.5

20

30

40

55

 

 

Voltage (V dc)

 

 

 

Communication requires a minimum loop resistance of 250 ohms.

Temperature Limits

Ambient

 

0 to 185

°F (-18 to 85 °C)

 

with meter option:

 

0 to 175

°F (-18 to 80 °C)

 

Process(1) (2) (3)

 

0 to 250

°F (-18 to 121 °C) Coplanarflange

 

0 to 300

°F (-18 to 149 °C) traditional flange

Storage

0 to 212 °F (-18 to 100 °C) with meter option:

0 to 185 °F (-18 to 85 °C)

(1)Process temperatures above 185 °F (85 °C) require derating the ambient temperature limits by 1.5 °F per degree above 185 °F.

(2)220 °F (104 °C) limit in vacuum service; 130 °F (54 °C) for pressures below 0.5 psia

(3)EPR O-ring is limited to 150 °F (66 °C) process temperature.

5-5