FIGURE 3-1. Typical

Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

FIGURE 3-1. Typical

Installation Flowchart.

START HERE

Bench

Calibration? No

Yes

Configure (Section 2)

Set Units

Set Range

Points

Set Output

Type

Set Damping

Verify

Apply Pressure

Within

Specifica- Yes

tions?

No

Refer to

Section 4

Troubleshooting

Field Install

Check Jumpers

and Switches

(page 2-2)

Mount

Transmitter

(pages 3-4–3-14)

Wire Transmitter (pages 3-12–3-14)

Power

Transmitter

(page 3-12)

Check Process

Connection

for Leaks

(page 3-12)

Confirm

Transmitter

Configuration

(page A-4)

Trim Transmitter

for Mounting

Effects

(page 2-15)

Done

3-3