Custom Meter

Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Custom Meter

Configuration

HART Comm

1, 3, 7, 2

 

 

The meter features a two-line display with five digits for reporting the process variable on the top line and six characters for displaying engineering units on the bottom line. The LCD meter can also display flow and level scales. The meter uses both lines to display diagnostic messages. The meter can be configured to display the following information:

engineering units

percent of range

user-configurable LCD scale

alternating between any two of the above

The user-configurable scale is a feature that enables the LCD meter to display flow, level, or custom pressure units. The meter can be configured using a HART Communicator (see Table A-1 on page A-4).

The user-configurable scale feature can define:

decimal point position

upper range values

lower range values

engineering units

transfer function

To configure the meter with a HART communicator, perform the following procedure:

1. Connect the communicator to the transmitter. Before connecting a communicator in an explosive atmosphere, make sure the instruments in the loop are installed according to intrinsically safe or nonincendive field wiring practices.

2.From the ONLINE screen, select 1 Device Setup, 3 Basic Setup, 7 Meter Options, 2 Custom Meter Setup.

3.To specify decimal point position:

a.Select 1 Sel dec pt pos. Choose the decimal point representation that will provide the most accurate output for your application. For example, when outputting between 0 and 75 GPM, choose XX.XXX.

b.Go to Step 8.

4.To specify a custom upper range value:

a.Select 2 CM Upper Value. Type the value that you want the transmitter to read at the 20 mA point.

b.Go to Step 8.

5.To specify a custom lower range value:

a.Select 3 CM Lower Value. Type the value that you want the transmitter to read at the 4 mA point.

b.Go to Step 8.

6.To define custom units:

a.Select 4 CM Units. Enter the custom units (five characters maximum) that you want the meter to display.

b.Go to Step 8.

7.To choose the transmitter transfer function for the meter:

See “Safety Messages” in Section 6 for warning information.

6-2