Emerson 3051N manual Overpressure Effect Static Pressure Effect, Power Supply Effect Load Effect

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 54
Image 54
Overpressure Effect

Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Overpressure Effect

Static Pressure Effect

Maximum zero shift after overpressure equal to Maximum Working Pressure

Rosemount 3051ND

Range Code

Overpressure Effect

 

 

 

 

 

0

– 3

± 0.5% URL

 

 

 

 

 

4

– 5

± 3.0% URL

 

 

 

 

Rosemount 3051NG

 

 

 

 

Range Code

Overpressure Effect

 

 

 

 

 

2

– 4

± 0.25% URL

 

 

 

 

 

 

5

± 0.30% URL

 

 

 

 

Rosemount 3051NA

 

 

 

 

Range Code

Overpressure Effect

 

 

 

 

 

0

– 4

± 0.05% URL

 

 

 

 

 

Rosemount 3051ND

Zero Error (can be calibrated out at line pressure) Per 1000 psi (6,9 MPa) line pressure

Range Code

Static Pressure Zero Effect(1)

0(2)

± 0.125% URL for Ps 750 psi (5,2 MPa)

1

± 0.25% URL for Ps 2000 psi (13,8 MPa)

 

 

2, 3

± 0.05% URL for Ps 2000 psi (13,8 MPa)

 

± [0.1 + 0.1 (Ps-2000) / 1000]% URL for

 

Ps > 2000 psi 3626 psi

 

 

4, 5

± 0.1% URL for Ps 2000 psi (13,8 MPa)

±[0.2 + 0.2 (Ps-2000) / 1000]% URL for Ps > 2000 psi 3626 psi

(1)Ps equals static line pressure applied

(2)Specification for Rosemount 3051N Range 0 is expressed in [% per 100 psi (689 KPa)]

Rosemount 3051ND

Span Error

Per 1000 psi (6,9 MPa) line pressure

Range Code

Static Pressure Span Effect

 

 

0(1)

± 0.15% input reading

1

± 0.40% input reading

 

 

2, 3

± 0.10% input reading

 

 

4, 5

± 0.20% input reading (uncertainty after calibration correction for

 

systematic effects)

 

 

(1)Specification for Rosemount 3051N Range 0 is expressed in [% per 100 psi (689 KPa)] up to 750 psi (5 171 KPa)

Power Supply Effect

Load Effect

Mounting Position Effect

Less than ±0.005% of calibrated span per volt for RDF ≤ 10

No load effect other than change in voltage supplied to the transmitter

Rosemount 3051ND/NG

Zero shifts up to ± 1.25 in H20 (0,31 KPa), which can be calibrated out; no span effect

Rosemount 3051NA

Zero shifts up to 2.5 in H20 (63,5 mm), which can be calibrated out; no span effect

5-4

Page 54
Image 54
Emerson 3051N manual Overpressure Effect Static Pressure Effect, Power Supply Effect Load Effect, Mounting Position Effect