Emerson 3051N manual Model Range Code Hydrostatic Test Pressure

Models: 3051N

1 100
Download 100 pages 43.51 Kb
Page 52
Image 52

Rosemount 3051N

Reference Manual

00809-0100-4808, Rev CA June 2008

Environmental

Quality Assurance Program

Nuclear Cleaning

Hydrostatic Testing

Performance to normal operating limits as described in the “Performance Specifications” and “Functional Specifications” sections of this manual.

In accordance with 10CFR50 Appendix B, ISO 9001:2000

To <1 ppm chloride content

Model

Range Code

Hydrostatic Test Pressure(1)

 

 

 

3051ND

0

750 psi

 

1

2000 psi

 

2–5

4200 psi

 

 

 

3051NG

2–5

150% of maximum working pressure(2)

3051NA(3)

1–4

 

(1)Process O-ring Code A (glass-filled TFE).

(2)Maximum Working Pressure equals upper range limit (URL).

(3)Hydrostatic testing not performed on Rosemount 3051NA Range Code 0.

5-2

Page 52
Image 52
Emerson 3051N manual Model Range Code Hydrostatic Test Pressure