35040

Operators Manual

3.9.8 Using the DETECTOR SELECT Button

Pressing the DETECTOR SELECT button displays the Detector Select screen, see Figure 3-18. This screen displays information about the selected ion chamber calibration factor. An ion chamber consists of two text lines and an ion chamber calibration factor (ICCF) with units. All ion chambers are entered and can be modified via the Customization Software.

The “f” cursor points to the first description string of the selected chamber. Use the UP arrow to select the next ion chamber and the DOWN arrow to select the previous ion chamber.

ADC

Farmer - S/N 12345

Co-60

ICCF=4.678E+09 R/C

Figure 3-18. The Detector Select Screen

The ion chamber calibration unit (R/C shown here) includes the radiological measurement unit (R). This is the radiological unit that appears on the Measurement screens.

NOTE

The ion chamber’s calibration unit includes the radiological measurement unit.

If you want radiological units that are not available, use the Customization software (see Section 3.10) to enter a new ICCF with the appropriate units.

The descriptive text should contain enough information to let the operator correctly select an ICCF that is appropriate for the measurement desired, e.g. energy, modality, build up, etc.

3.9.9 Using the AIR DENSITY Button

Pressing the AIR DENSITY button displays the Air Density screen, see Figures 3-19 and 3-20. This screen contains the temperature value, pressure value, and Air Density Correction Factor (ADCF). Press the AIR DENSITY button to advance the “f” cursor to the next field. To change the value or state of an item at the “f” cursor, use the UP and DOWN arrows. The Model 35040 Therapy Dosimeter displays the Air Density Correction (ADC) annunciator in the upper left-hand corner of all screens if the air density correction is enabled, see Figure 3-19.

ADC

760 mmHg

22.5 C

ADCF = 1.002

Figure 3-19. The Air Density Screen With Air Density Corrections Enabled

To disable the air density correction, advance the “f” cursor to the ADC annunciator at the upper left of the display. With the “f” cursor at this position, pressing the UP and DOWN keys will toggle the air density correction on (ADC) and off (OFF).

3-14

Page 38
Image 38
Fluke 35040 manual Using the Detector Select Button, Using the AIR Density Button