Fluke 35040 manual Measurements. Return, Symptom Current Overrange Displayed

Models: 35040

1 94
Download 94 pages 19.16 Kb
Page 62
Image 62

35040

Operators Manual

7

Additional operator settings

Select the proper settings

are invalid.

from the front panel.

 

8

Timer interval customization

Reinstall settings using the

settings are invalid.

Customization software.

 

 

Additional calibration

Do NOT make

9

settings are invalid.

measurements. Return

 

 

instrument for repair.

Explanation:

The instrument detects whether the stored parameters are valid at power-up. If the

 

parameters are invalid, it will affect one or more of the measurements. The operator may

 

restore some of the information either by selecting options from the front panel or by

 

running the customization software. Invalid calibration settings require a new calibration.

 

Follow the recommendations in the above table.

Symptom:

"Current Overrange"

displayed.

ADC

 

 

Current Overrange

 

Reduce exposure and

 

Press RESET/MEASURE

Cause:

The input current has exceeded the 1.200 uA limit.

Action:

Stop the exposure. Change the setup to reduce the current.

Explanation:

The instrument has a 1.000 uA limit on measurable current and will continue to measure

 

charge for currents up to 1.200 uA. The instrument can accept an input current up to 3

 

mA without damage to the instrument.

Symptom:

“Current Overrange” displayed.

 

ADC

 

Current Overrange

 

Reduce exposure and

 

Press RESET/MEASURE

Cause:

The input current has exceeded the 100 nA limit.

Action:

Stop the exposure. Change the setup to reduce the current.

Explanation:

The instrument has a 100 nA limit on measurable current. The instrument can accept an

 

input current up to 3 mA without damage to the instrument.

Symptom:

Dashes displayed instead of digits.

Cause:

No meaningful measurement could be obtained.

Action:

Press the RESET/MEASURE button to reset the measurements.

Explanation:

ICBias —.- V : the bias voltage measurement is unavailable due to invalid signals within

 

the instrument. This is a normal momentary condition when the instrument is powered

 

up.

Symptom:

The bias voltage measurement changes by one digit.

4-4

Page 62
Image 62
Fluke 35040 manual Measurements. Return, Symptom Current Overrange Displayed, Symptom Dashes displayed instead of digits