Fluke 35040 manual Units Display Minimum Maximum, Standard Resolution Error

Models: 35040

1 94
Download 94 pages 19.16 Kb
Page 24
Image 24

35040

Operators Manual

Table 2-2.

Air Density Adjustment Ranges and Resolutions

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Units

Display

Minimum

Maximum

Standard

Resolution

% Error*

 

inches of mercury

inHg

20.55

31.30

29.90

0.05

0.12

 

 

 

 

 

 

 

 

 

millimeters of mercury

mmHg

522

795

760

1.0

0.10

 

 

 

 

 

 

 

 

 

altitude in feet

ft

-1200

10,000

0

25

0.07

 

 

 

 

 

 

 

 

 

altitude in meters

m

-360

3050

0

10

0.09

 

 

 

 

 

 

 

 

 

millibars

mbar

697

1060

1013

1.0

0.07

 

 

 

 

 

 

 

 

 

hectoPascals

hPa

697

1060

1013

1.0

0.07

 

 

 

 

 

 

 

 

 

Units

Display

Minimum

Maximum

 

 

 

 

Fahrenheit

F

32.0

122.0

 

 

 

 

Kelvin

K

273.0

323.0

 

 

 

 

Celsius

C

0.0

50.0

 

 

 

 

Standard

Resolution

% Error*

71.6

68.0

0.5

0.05

 

 

 

 

295.2

293.2

0.5

0.09

 

 

 

 

22.0

20.0

0.5

0.09

 

 

 

 

*The “% Error” column indicates the maximum amount of error that can occur due to the step resolution of the pressure or temperature units.