35040

Operators Manual

4.5 Bias Error

 

 

 

Go to

 

 

 

line:

1.

Is "??Bias" displayed?

YES

2

NO

6

 

 

2.

Has the instrument been on for more than one minute?

YES

3

NO

8

 

 

3.

Turn off the power. Wait 10 seconds. Turn on the power. Wait 1 minute. Is

YES

4

"??Bias" displayed?

NO

--

 

4.

PROBLEM: Electrometer fault. See Section 4.10.

YES

5

NO

--

 

 

5.

Change the bias voltage to a different setting and then change back. Wait

 

 

about 10 seconds. Is the error on the display?

 

 

 

 

 

 

Change the bias voltage to 0.0 V. Disconnect all triax cables. Cap the

YES

7

6.

connectors. Change bias voltage to desired setting. Wait about 10 seconds.

NO

8

 

Is the error on the display?

 

 

 

7.

Change the bias voltage to 0.0 V. Examine the triax connectors on the

 

 

instrument, both front and rear, for foreign matter. Is foreign matter present?

 

 

 

 

 

8.

See Section 4.7.

YES

9

NO

18

 

 

9.

Normal operation established.

YES

10

NO

11

 

 

10.

Do you use a separate cable to connect the ion chamber to the instrument?

 

 

 

 

 

 

 

Change the bias voltage to 0.0 V. Connect only the cable to the instrument.

YES

12

11.

Change the bias voltage to the desired setting. Wait about 10 seconds. Is the

NO

13

 

error on the display?

 

 

 

12.

PROBLEM: Bias fault in cable. See Section 4.10.

 

 

 

 

 

 

 

Change the bias voltage to 0.0 V. Connect the ion chamber. Change the bias

YES

14

13.

voltage to the desired setting. Wait about 10 seconds. Is the error on the

NO

16

 

display?

 

 

 

14.

PROBLEM: Bias fault in instrument. See Section 4.10.

YES

15

NO

17

 

 

15.

PROBLEM: Bias fault in ion chamber. See Section 4.10.

 

 

 

 

 

 

16.

This message is displayed until the signals from the isolated side are valid.

 

 

 

 

 

 

4-8

Page 66
Image 66
Fluke 35040 manual Bias Error, Go to, Line