35040

Operators Manual

ADC LoBias Tmr ALC 39.995 nC

659.8 pA

59.891 s

Figure 3-32. Annunciator positions

Table 3-2 lists the indicators the instrument uses to signal that a measurement quantity has not been compensated for leakage or is not available due to the specified condition.

Table 3-2. Measurement Indicators and Conditions

Indicator

Measurement

Condition

 

 

Automatic leakage compensation is enabled, but

? (left of value)

All Non-leakage1

the exposure started before a leakage

 

 

measurement was available.

 

 

Automatic leakage compensation is enabled, but

-.---

Leakage

the exposure started before a leakage

 

 

measurement was available.

 

 

The calculated effective time is longer than the

INVALID

Effective Time

actual exposure. This condition occurs when a

large amount of charge is injected at the

 

 

 

 

beginning of the exposure.

NA

Leakage

The leakage measurement is not available since

automatic leakage compensation is disabled.

 

 

 

 

The quantity or one of its dependents has

OVER

Non-leakage1

exceeded the maximum value. Refer to Table 3-

 

 

3 for more information.

1All non-leakage measurements (e.g., charge, current, dose, rate, effective time, etc.)

Table 3-3 specifies the conditions that cause the Model 35040 Therapy Dosimeter to display an overrange indicator (OVER) for a measurement quantity.

Table 3-3. Overrange Conditions

Condition

 

Overranged Measurement1

 

charge

charge > 19.999 mC

average current

accumulated charge

 

 

effective exposure time

accumulated charge > 1.0000 C

accumulated charge

current > 1.000 μA

current

current > 1.200 μA

all non-leakage

time > 19,999.9 s

effective exposure time

1Includes radiological measurements derived from electrical measurements.