35040

Operators Manual

ADC

OVER nC OVER pA

OVER min OVER s

ADC

Current Overrange Reduce exposure and Press RESET/MEASURE

Figure 3-28. The Measurement Screen With the Charge and Current Limits Exceeded After the First 60 Seconds

Figure 3-29. The Measurement Overrange Screen

Table 3-1 specifies the conditions that cause the Model 35040 Therapy Dosimeter to display the overrange indicator for a measurement.

Table 3-1. Overrange Conditions

Condition

Values Overranged

 

 

charge > 2.4 μC

charge, accumulated charge, dose, accumulated dose, average

 

current, average dose rate, effective exposure time

accumulated charge > 1.0000 C

accumulated charge, accumulated dose

 

 

current > 100.0 nA

all except leakage

 

 

time > 59 min 59 s

effective exposure time

 

 

3.9.14 Reset Error

Pressing the RESET /MEASURE button while on the Measurement Screen causes the Model 35040 Therapy Dosimeter to reset the electrometer, zero all readings, and restart the measurement process. If an exposure starts during the reset operation, the instrument displays the reset error message (see Figure 3-30).

ADC

Internal Reset Error

Press RESET/MEASURE

and Repeat Exposure

Figure 3-30. The Measurement Screen Showing a Reset Error

To remove the error condition, end the exposure, and press the RESET/MEASURE button. Wait a few seconds and repeat the exposure.

3-18

Page 42
Image 42
Fluke 35040 manual Reset Error, Overrange Conditions, Condition Values Overranged