Troubleshooting and Service

Troubleshooting 4

Explanation: The low battery annunciator is a reminder that while running on the battery, there is a finite amount of energy available. For the battery in this instrument and the energy consumption of the instrument, when the battery voltage drops below 5.9 volts there is less than 30 minutes of operating time left.

Symptom:

“Internal Reset Error” displayed.

 

ADC

 

Internal Reset Error

 

Press RESET/MEASURE

 

and Repeat Exposure

Cause:

Start of exposure detected during reset process.

Action:

Stop the exposure. Press the RESET/MEASURE button. Wait for the “System Not RDY”

 

annunciator to extinguish. Start the exposure.

Explanation: The instrument measured a current above the start threshold while resetting the electrometer. It is assumed that the exposure was started while the reset was occurring. If the exposure did start during the reset, some unknown amount of charge was lost. The lost charge would lead to an incorrect reading.

Symptom:

“Configuration Error” displayed.

ADC

Configuration Error

Service Required

Error Number : 16

Cause:

The instrument has lost the calibration or customization data.

Action:

Convert the configuration error number to a binary value using a program such as the

 

Windows® Calculator and use the table below to determine the extent of the problem.

 

For example, a configuration error of “18” corresponds to a binary value of “10010” (bits 5

 

and 2 set), which indicates that the customization for both the ion chamber and settings

 

are invalid.

 

 

 

 

 

 

 

 

 

Bit

Description

Action

 

 

 

 

 

 

 

1

Operator settings are

Select the proper settings

 

 

invalid.

from the front panel.

 

 

 

 

 

2

Customization settings are

Reinstall settings using the

 

 

invalid.

Customization software.

 

 

 

 

 

 

Electrometer calibration

Do NOT make

 

 

3

settings are invalid.

measurements. Return

 

 

 

 

instrument for repair.

 

 

 

Bias calibration settings are

Do NOT make

 

 

4

invalid.

measurements. Return

 

 

 

 

instrument for repair.

 

 

5

Ion chamber customization

Reinstall ion chambers using

 

 

settings are invalid.

the Customization software.

 

 

 

 

 

6

Ion chambers cannot be

Return instrument for repair

 

 

read.

to correct problem.

 

 

 

 