B.4.6 Annunciators, Indicators, and Overrange Conditions

Table B-1 specifies the annunciators that may appear on the top line of the display.

Table B-1. Annunciators

Annunciator

Condition

 

 

??Bias

No bias measurement possible since the

electrometer is not ready

 

ADC

Air density correction in use (only applicable to

radiological measurements)

 

ALC

Automatic leakage compensation enabled

 

 

LoBat

Low Battery voltage

 

 

LoBias

Low Bias voltage

 

 

HiBias

High Bias voltage

 

 

OvAmps

Overranged current

 

 

Rst

Resetting electrometer

 

 

Tmr

Timer in use

 

 

WAIT Bias Not Ready

The dosimeter is setting the bias

 

 

WAIT System Not Rdy

The dosimeter is resetting the electrometer

 

(only applicable when automatic leakage

 

compensation is enabled)

Figure B-8 illustrates the locations of some annunciators. The battery, bias, and overranged current annunciators appear at the same location, and the instrument alternates the displaying of each annunciator when more than one is applicable.

ADC LoBias Tmr ALC 39.995 nC

659.8 pA

59.891 s

Figure B-8. Annunciator Positions

Table B-2 lists the indicators the instrument uses to signal that a measurement quantity has not been compensated for leakage or is not available due to the specified condition.

Table B-2. Measurement Indicators and Conditions

Indicator

Measurement

Condition

 

 

 

? (left of value)

All Non-leakage1

Automatic leakage compensation is enabled, but the exposure

 

 

started before a leakage measurement was available.

-.---

Leakage

Automatic leakage compensation is enabled, but the exposure

 

 

started before a leakage measurement was available.

INVALID

Effective Time

The calculated effective time is longer than the actual

 

 

exposure. This condition occurs when a large amount of

 

 

charge is injected at the beginning of the exposure.

NA

Leakage

The leakage measurement is not available since automatic

 

 

leakage compensation is disabled.

OVER

Non-leakage1

The quantity or one of its dependents has exceeded the

 

 

maximum value. Refer to Table B-3 for more information.