35040

Operators Manual

With the automatic leakage compensation settings displayed (Figure 3-14), press the TEST FUNCTION button. The "f" cursor will appear to the left of the ALC annunciator. Use the UP and DOWN arrow buttons to toggle the setting between on (ALC) and off (OFF).

With automatic leakage compensation enabled, press the TEST FUNCTION button to advance the "f" cursor to the next field or exit Edit mode. When the cursor points to a threshold setting, use the UP and DOWN arrow buttons to increment and decrement the value, respectively. Nominal values for the start and stop thresholds are 1.3 and 1.2 pA, respectively. However, depending on the system leakage, other values may be necessary.

When automatic leakage compensation is disabled, the annunciator displays off (OFF) (Figure 3-15), the start and stop threshold settings cannot be changed and display Not Available (NA), as do the leakage measurements.

ADC

 

ALC

Start Thres:

1.0 pA

Stop

Thres:

0.9 pA

Current:

0.0 pA

Figure 3-14.

Automatic Leakage Compensation Enabled

ADC

 

OFF

Start Thres:

NA pA

Stop

Thres:

NA pA

Current:

0.0 pA

Figure 3-15. Automatic Leakage Compensation Disabled

NOTE

The instrument may reset all measurements when the automatic leakage compensation parameters are

changed.

NOTE

Disabling automatic leakage compensation prevents automatic exposure detection and the measurement of leakage current, and leakage measurements will display Not Available (NA).

Figure 3-16 shows the name of the calibration agency and the date the instrument was calibrated. The bottom three lines may be different on your instrument depending on what the calibration agency puts here.

ADC

Calibrated by

Keithley RMD

Nov 15, 1993

Figure 3-16. The Test Function Screen 5 - Calibration Identification