35040

Operators Manual

4.3 Display Will Not Light

 

 

 

Go to

 

 

 

line:

1.

Is the instrument connected to line voltage?

YES

2

NO

4

 

 

2.

Disconnect the line cord. Check the fuse. (See Section 4.6.) Is the fuse

YES

3

filament intact?

NO

11

 

3.

Attach known operational appliance to outlet. Does the appliance work?

YES

4

NO

12

 

 

4.

Connect instrument to AC line. Press the Power button.

YES

5

Did the display light up?

NO

7

 

5.

Is the battery current between +0.10 A and +3.00 A?

YES

6

NO

13

 

 

 

The fault has been cleared. The battery is charging. Leave the instrument

 

 

6.

plugged in for at least 2 hours. When the battery is less than 5.4 volts and the

 

 

instrument is running on battery, the instrument will turn itself off. This is to

 

 

 

 

 

 

protect the battery from becoming severely discharged.

 

 

7.

Replace cord with known good cord or establish that the line cord is good. Is

YES

8

line cord good?

NO

14

 

8.

Was the battery low the last time the instrument was stored?

YES

9

NO

16

 

 

9.

Has the instrument been plugged in for more than 24 hours?

YES

10

NO

15

 

 

10.

PROBLEM: Instrument will not power up after 24 hours. See Section 4.10.

 

 

 

 

 

 

11.

See Section 4.6.

 

 

 

 

 

 

12.

Faulty Outlet. Use a different outlet.

 

 

 

 

 

 

13.

See Section 4.2 to isolate problem.

 

 

 

 

 

 

14.

ORDER: Proper line cord. See Section 4.9.

 

 

 

 

 

 

 

The battery may take more than 18 hours to recharge if the battery was in the

 

 

15.

low battery condition and was not recharged before storage. Leave the

 

 

 

instrument plugged in and try again in 18-24 hours.

 

 

16.

PROBLEM: Instrument will not power up. See Section 4.10.

 

 

 

 

 

 

4-6

Page 64
Image 64
Fluke 35040 manual Display Will Not Light