Fluke 35040 manual Setup, New Features B

Models: 35040

1 94
Download 94 pages 19.16 Kb
Page 79
Image 79

Appendix

New Features B

1.000 μA

0.001 μA

1

Average Current Measurements

Range and Resolution:

Full Scale Range

Display Resolution

 

 

200.00 pA

0.01 pA

 

 

2.0000 nA

0.0001 nA

 

 

20.000 nA

0.001 nA

 

 

200.00 nA

0.01 nA

 

 

1.0000 μA

0.0001 μA

 

 

Effective Exposure Time Measurements

Range and Resolution:

 

Full Scale Range

Display Resolution

 

 

 

Seconds

 

Hours/Minutes/Seconds

 

 

 

 

 

 

59.99 s

 

59.99 s

0.01 s

 

 

 

 

19,999.9 s

 

5 h 33 min 19.9 s

0.1 s

 

 

 

 

Measurement Method: The effective exposure time value is calculated as the quotient of the charge and average current measurement values. A minimum exposure duration of 3.5 seconds is required to produce an effective exposure time measurement value.

B.3 SETUP

B.3.1 Making Non-Timed Exposure Measurements

With Automatic Leakage Compensation Disabled

The instrument must be manually reset before each exposure when automatic leakage compensation is disabled.

Procedure

Hold RESET/MEASURE for one second just before making the exposure to reset the instrument. Note that the instrument displays the “Rst” annunciator while the hardware is being reset.

Wait for the reset to complete.

Make the exposure.

The exposure measurements will be displayed but may drift due to leakage. Repeat this sequence for every exposure.

With Automatic Leakage Compensation Enabled

In addition to compensating measurements for leakage, the instrument holds the exposure measurements and automatically resets after each exposure when automatic leakage compensation is enabled.

B-3

Page 79
Image 79
Fluke 35040 manual Setup, New Features B