Appendix

Operation Instructions B

the instrument will cease updating the exposure measurements when the current drops below the stop threshold and will return to measuring leakage.

NOTE

The instrument uses the start and stop threshold settings to sense when an exposure is in progress.

With the automatic leakage compensation settings displayed (Figure B-4), press the TEST FUNCTION

button. The "" cursor will appear to the left of the ALC annunciator. Use the UP and DOWN arrow buttons to toggle the setting between on (ALC) and off (OFF).

With automatic leakage compensation enabled, press the TEST FUNCTION button to advance the "" cursor to the next field or exit Edit mode. When the cursor points to a threshold setting, use the UP and DOWN arrow buttons to increment and decrement the value, respectively. Nominal values for the start and stop thresholds are 1.3 and 1.2 pA, respectively. However, depending on the system leakage, other values may be necessary.

When automatic leakage compensation is disabled, the annunciator displays off (OFF) (Figure B-5), the start and stop threshold settings cannot be changed and display Not Available (NA), as do the leakage

ADC

 

1.0

ALC

Start Thres:

pA

Stop

Thres:

0.9

pA

Current:

0.0

pA

measurements.

 

 

Figure B-4.

Automatic Leakage Compensation Enabled

Figure B-5.

Automatic Leakage Compensation Disabled

NOTE

The instrument may reset all measurements when

ADC

Thres:

 

OFF

Start

NA pA

Stop

Thres:

NA

pA

Current:

0.0

pA

the automatic leakage compensation parameters are changed.

NOTE

Disabling automatic leakage compensation prevents automatic exposure detection and the measurement of leakage current, and leakage measurements will display Not Available (NA).

B-7

Page 83
Image 83
Fluke 35040 manual Start NA pA Stop Thres Current, Operation Instructions B