8.3 Calibration concentration setting (except zirconia O2 sensor)

Set a span value for the calibration concentration setting. (Zero point calibration concentration is fixed at
zero.)
Set the calibration gas concentration (span value).
When FUNC key is pressed in the measuring sta-
tus, the previously set span value will appear on
the main indication.
The CAL SET LED of the function indicating
lamps will flicker.
By pressing the key, the highest digit of the
main indication for the 1st component will flicker,
and the span value can not be set.
Select a range with the keys.
After selecting the range, set a span value in this
status.
The numeric value will be incremented by press-
ing key.
The digit to be set can be selected by pressing
key.
For selecting the 2nd and subsequent compo-
nents and the range, press the key and
while the main indication for the 1st compo-
nent is flickering, press the COMP key and
then the main indication for the component to
be set will flicker. Now press the keys to
select the range to be set.
When the span value has been set, press the ENT
key. Setting operation is completed.