Example of auto calibration time chart for 4-component measurement

Auto calibration start time

 

 

 

 

 

 

Under

 

 

Under auto calibration

 

 

 

measurement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Zero

Span

Span

Zero

Span

Span

Span

Substi-

 

gas

gas 1

gas 3

gas

gas 1

gas 2

gas 3

tution

• Gas flow time

 

CO

O2

 

NO

SO2

O2

 

 

 

 

 

 

 

 

 

Auto calibration

T

T

T

T

T

T

T

T

gas flow time

 

 

 

 

 

T: settable from 100 to 599 sec. (Note)

Auto calibration cycle

Gases introduced:

• Sample gas

Zero gas

Span gas 1

CO

Span gas 1

NO

Span gas 2

SO2

Span gas 3

O2

Indicator lamps

AUTO CAL.

Flickers

ZERO

Flickers

SPAN

CAL

Lights

HOLD

(at HOLD ON)

Auto calibration abnormality:

If an abnormality occurs during auto calibration, a contact signal for auto calibration abnormality will be output from the input/output terminals on the upper of the analyzer. The gas which incurred the abnormality will not be calibrated, and instead the next component will be calibrated.

Note: Usually, T is set at 360 sec. for measurement including SO2, and at 180 sec. for other measurements. For change of setting, refer to the separate instruction manual of infrared gas analyzer.

– 19 –