Example of auto calibration time chart for   | ||||||||
Auto calibration start time  | 
  | 
  | 
  | 
  | 
  | 
  | ||
Under  | 
  | 
  | Under auto calibration  | 
  | 
  | 
  | ||
measurement  | 
  | 
  | 
  | 
  | 
  | |||
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
  | Zero  | Span  | Span  | Zero  | Span  | Span  | Span  | Substi-  | 
  | gas  | gas 1  | gas 3  | gas  | gas 1  | gas 2  | gas 3  | tution  | 
• Gas flow time  | 
  | CO  | O2  | 
  | NO  | SO2  | O2  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
Auto calibration  | T  | T  | T  | T  | T  | T  | T  | T  | 
gas flow time  | ||||||||
  | 
  | 
  | 
  | 
  | T: settable from 100 to 599 sec. (Note)  | |||
Auto calibration cycle
Gases introduced:
• Sample gas
Zero gas
Span gas 1
CO
Span gas 1
NO
Span gas 2
SO2
Span gas 3
O2
• Indicator lamps
AUTO CAL.
Flickers
ZERO
Flickers
SPAN
CAL
Lights
HOLD
(at HOLD ON)
Auto calibration abnormality:
If an abnormality occurs during auto calibration, a contact signal for auto calibration abnormality will be output from the input/output terminals on the upper of the analyzer. The gas which incurred the abnormality will not be calibrated, and instead the next component will be calibrated.
Note: Usually, T is set at 360 sec. for measurement including SO2, and at 180 sec. for other measurements. For change of setting, refer to the separate instruction manual of infrared gas analyzer.