4. INSTALLATION

Danger

This product is not explosion-proof. Do not use it in explosive gas-rich atmosphere. Otherwise, serious accidents such as explosion and fire may be caused.

Caution

Installation, carrying and relocation must be made by a qualified contractor or your dealer. Incorrect installation results in falling accidents, electric shocks or fire accidents.

The gas analyzer is heavy. Special care should be taken for installation to prevent trouble or injury due to falling or dropping of the analyzer.

When the gas analyzer needs to be lifted, be sure to put on gloves. Do not hold it with bare hands to prevent injury.

This product must be installed in a place that meets the conditions specified in the ‘instruction manual.’ Use in a place that is outside the given installation conditions may cause an electric shock, fire or mal- function.

During installation work, attention should be paid not to allow entrance of wire fragments, etc. into the product. Otherwise, a fire, trouble or malfunction may be caused.

There are a number of conditions which must be met for operating the gas analyzer stably over a long time.

4.1Selection of installation site

A site which satisfies the following conditions should be selected for installation of the analyzer.

(1)A place which is near the gas sampling location

(2)A place which is not exposed to direct sunlight or heat radiated from a high temperature object (One which is not exposed to rain or water splash in case of indoor type)

(3)A place which has a clean atmosphere (dust and corrosive gas not present)

(4)A place with a minimum of vibration

(5)A place where daily variation in temperature is minimum

4.2Selection of gas sampling location

A place which satisfies the following conditions should be selected.

(1)A typical place where the gases to be analyzed exist

(2)Where the gases always flow without stagnation

(3)A place easy of access for maintenance and inspection

(4)Where the gases respond quickly to a change in the process operating conditions

(5)Where variation in pressure, temperature, etc. is minimum