Trouble

Probable location

Check item

Remedy

 

 

 

 

 

Gas leakage

Check for gas leakage before the

Retighten or replace parts.

Indicated value is

not as expected.

 

aspirator.

 

 

Interference due to

With respect to zero gas indication, the

Readjust to compensate for

 

moisture

indicated value when air is introduced

moisture interference.

 

 

exceeds ± 2% full scale.

 

 

 

 

 

 

Measuring range

Check if correct range is set.

Reset the range.

 

 

 

 

 

Zero, span

Check zero and span using standard gas.

Adjust for correct zero

 

 

 

point and span.

 

 

 

 

 

Cell window

Check if cloudy

Clean the window.

 

 

 

 

 

Power supply, fuse

Check the power supply voltage and fuse.

Replace the fuse.

Indication doesn’t

change

 

 

 

 

 

 

 

13.2Error codes and remedies

This analyzer is provided with self-diagnosis functions, and an error code is displayed if an abnormality occurs in the instrument.

Take a proper measure with reference to the instruction manual of infrared gas analyzer when an error code appears.

– 44 –

Page 52
Image 52
Fuji Labs ZSP instruction manual Error codes and remedies