Trouble | Probable location | Check item | Remedy |
|
|
|
|
| Gas leakage | Check for gas leakage before the | Retighten or replace parts. |
Indicated value is | |||
not as expected. |
| aspirator. |
|
| Interference due to | With respect to zero gas indication, the | Readjust to compensate for |
| moisture | indicated value when air is introduced | moisture interference. |
|
| exceeds ± 2% full scale. |
|
|
|
|
|
| Measuring range | Check if correct range is set. | Reset the range. |
|
|
|
|
| Zero, span | Check zero and span using standard gas. | Adjust for correct zero |
|
|
| point and span. |
|
|
|
|
| Cell window | Check if cloudy | Clean the window. |
|
|
|
|
| Power supply, fuse | Check the power supply voltage and fuse. | Replace the fuse. |
Indication doesn’t | |||
change |
|
|
|
|
|
|
|
13.2Error codes and remedies
This analyzer is provided with
Take a proper measure with reference to the instruction manual of infrared gas analyzer when an error code appears.
– 44 –