Automatic calibration

Calibration tolerance:

Automatic calibration output hold:

±0.2% FS Possible

(Example in 4 component analyzer):

CO zero

CO span

O2 zero

NOx, SO2 zero

NOx span

SO2 span

O2 span

Automatic calibration

• Calibration cycle:

Once/day (variable, up to 7 days)

Calibration gas feed time:2 to 10 min. (variable), common to each gas

• Display:

Calibration failure

 

Error code (E-) is displayed when one-time calibration volume is more

 

than ± 20% FS on both zero and span and it is under calibration:

 

 

 

 

CAL

lamp ON

Ordering information

1.Measuring range of analyzer

2.Length of sampling pipe

3.Insertion length of sampling tube

4.Power voltage/frequency

5.Instruction required when standard gas is not necessary

6.Instruction required when individual inspection is not necessary

7.Necessity of spares and supplies (for 1 year)

8.Additional functions

Note: When special functions such as remote range selection, output hold by external contact, auto- matic calibration, automatic calibration start by external contact, insulation output, measured value concentration alarm, etc. are required, they should be separately designated.

– 48 –

Page 56
Image 56
Fuji Labs ZSP instruction manual Ordering information, Calibration tolerance Automatic calibration output hold, Display