Fuji Labs ZSP instruction manual Operation, Operating procedure

Models: ZSP

1 62
Download 62 pages 56.89 Kb
Page 33
Image 33

9. OPERATION

Danger

If you notice malodor or abnormal sound, stop the instrument immediately. If this is neglected, fire may occur.

Caution

When the analyzer is to be operated after it has been left unused for a long time, the operating procedure is different from the normal stop/start procedure. It should be handled according to the procedure shown in the instruction manual to ensure its proper performance and to prevent trouble or injury.

Do not operate the analyzer for a long time with its door left open to avoid entry of dust into the analyzer. Dust deposits will result in damage to the analyzer.

Prohibition

Do not insert a rod or your finger into the fan (electronic cooler at ceiling), otherwise injury may be caused by the rotating blades.

Do not touch metal, finger, etc. to the input and output terminal block of analyzer. Otherwise, shock hazard or injury may be caused.

Smoking and use of fire/flame is prohibited in the vicinity of the gas analyzer. Otherwise, a fire may occur.

A water-filled container such as a bucket should not be placed on the roof of the gas analyzer having indoor specifications. If such a container turns over, spilled water may cause shock hazard or burning inside the analyzer.

9.1Operating procedure

(1)Check if all the preparation for operation have been completed.

(2)Press the range selector switch on the indication/operation panel to select a range to be used.

(3)Press the function key “FUNC” on the indication/operation panel to light up the “MEAS” function indicator lamp.

(4)Turn on the aspirator power supply and confirm that the ball of flow meter is at 0.4N/min. Prescribed flow rate may be 0.5N/min. So it should be checked in the piping system diagram sepa- rately presented.

– 25 –

Page 33
Image 33
Fuji Labs ZSP instruction manual Operation, Operating procedure