[When adjusting for each range]

(Simultaneous calibration of H and L ranges has been factory-set).

Perform setting in the following way.

NO

 

By pressing

CAL

key while the highest digit of the

 

main indication for the 1st component is flickering,

 

then ‘

’ will appear on the main indication and

 

 

‘ ’ will appear on the sub indication.

NO

 

 

 

 

CAL

Press key and will change to .

= Meaning of set values =

: The calibration is valid only for the selected range, and zero and span calibration can

be made independently for each range.

: By conducting zero and span calibration

for one range, the calibration will be done

NO

NO

ENT

automatically for the other ranges as well.

When the selection has been finished, press the ENT key. Setting operation is completed.

– 13 –