8.5Zero calibration (air point calibration in case of zirconia O2 sensor)

This is used for adjusting the zero point.

Flow the zero gas (0.3N/min.) by adjusting the needle valve of pressure regulator and wait until the indica- tion stabilizes.

When the Zirconia O2 sensor is used, use air.Note)

When the indication has stabilized, select the measur-

ing range to be calibrated by using the keys.

When using a multi-component analyzer, all the components undergo zero calibration simulta- neously at the selected range.

Press the ZERO key, and ZERO key lamp will flicker.

Next press the CAL key and conduct zero calibra- tion.

The CAL key lamp lights steadily during calibra- tion.

NO

ZERO

NO

Flickers

ENT ZERO SPAN CAL

CAL

NO

Lights

ENT ZERO SPAN CAL

ENT ZERO SPAN CAL

( Lamp is turned off after calibration )

Since the measuring status is resumed after the cali- bration, carry out the calibration for other ranges if necessary.

Note: When using different gases for the low and high ranges, apply 9 to 10 Vol. % O2/N2 which has been set for the low range.

When the function for automatic calibration is provided, the solenoid valve turns on and cali- bration gas flows upon pressing the ZERO key.