Standard requirements

 

 

 

 

for sample gas:

Temperature

60 to 800° C

 

Dust

— 100 mg/Nm3 or less

 

Pressure

-2.94 to +2.94kPa

 

Components

SO2

0 to 1000ppm

 

 

 

NOx

0 to 2000ppm (NO2 100ppm or less)

 

 

 

CO2

several % to over 10%

 

 

 

CO

0 to 2000ppm

 

 

 

O2

0.5 to 21%

 

 

 

H2O

0 to 40%

 

 

 

HCN

0 to 100ppm

 

 

 

N2

remaining percent

When other components are contained in the sample gas, consult with Fuji.

Note) NO2/NO converter catalyst life: 1 year/NO2, 10ppm

1month/NO2, 100ppm

Installation requirements: Select a location free from direct sunlight and severe vibration.

Table 1 Number of output points and kinds of outputs

Type

ZSP1L

ZSP1F

ZSP1P

ZSP1B

ZSP1A

ZSP1H

ZSP1M

 

 

 

 

 

 

 

 

Measuring

NOx, SO2,

NOx, SO2,

NOx, O2

CO, O2

SO2, (O2)

NOx, CO2, O2

NOx, SO2,

component

CO, O2

O2

 

 

 

 

CO, CO2, O2

No. of

11

7

3

3

3

5

12

output points

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Kind of output

A, B, C,

A, B, D, E,

A, D, E

C, D, K

B, (D)

A, B, D, E, K

A, B, C,

 

D (2 points), E,

F, H, J

 

 

 

 

D (2 points),

 

F, G, H, J, K

 

 

 

 

 

E, F, G, H, J,

 

 

 

 

 

 

 

K, L

 

 

 

 

 

 

 

 

A: NOx instantaneous value

 

B: SO2 instantaneous value

 

C: CO instantaneous value

 

D: O2 instantaneous value

 

E: NOx instantaneous value after O2 correction

F: SO2 instantaneous value after O2 correction

G: CO instantaneous value after O2 correction

H: 1hr. moving average NOx value after O2 correction

J: 1hr. moving average SO2 value after O2 correction

K: 4hr. moving average CO value after O2 correction

L: O2 instanteneous value

 

 

 

 

 