SO2
CO
CO2
O2
Repeatability:
Drift:
Linearity: Response time:
Output:
Contact output:
Indication:
Specifications
Type of cubicle: Measuring system:
Sampling system:
Measuring range: (Range selection: Max. rate 1:20 (except for O2))

14. SPECIFICATIONS

Outdoor, self-standing, sealed type or indoor type

Non-dispersion infrared absorption (NDIR method) for NOx, SO2, CO and CO2; zirconia method for O2.

Dry sampling electronic cooling & dehumidification system (Permapure dryer adopted)

NOx — 0 to 100 .... 2000 ppm

— 0 to 100 .... 1000 ppm

— 0 to 200 .... 2000 ppm

— 0 to 5 ........ 50%

— 0 to 10/25%

Note: Measuring range of each gas is 50 to 5000ppm, available in approved type (special order).

± 0.5% of full scale

Zero — ± 2% of full scale/week for NOx, SO2 and CO

±2% of full scale/month for O2

Span — ± 2% of full scale/week

± 2% of full scale

90% response from the inlet of cubicle SO2 : 4 min. CO : 2.5 min NO : 2 min Others : 3 min.

4 to 20mA DC (allowable load resistance 550), non-isolated. For the number of output points, see Table 1.

Each NO (1a) contact (100V AC, 1A) for error (analyzer error, calibration error) and auto calibration status

Digital indication (indicating measured value and measuring range on analyzer front panel of infrared gas analyzer)

Sample gas extracting rate: Approx. 2 N/min.

Gas extractor:

Electrical heating type (filter built in)

 

Filter mesh — 40m mesh of 316 stainless steel

 

Probe

— 316 stainless steel, length 600, 800 or 1000mm

 

Flange

— JIS 5 K65AFF

Sample inlet tube:

ø 10/ ø 8mm teflon tube for CO and O2 meter and NOx and O2 meter

 

Heating tube ( ø 10/ ø 8mm teflon tube) for NOx, SO2 and O2 meter and NOx, SO2, CO

 

and O2 meter

Standard gas (3.4 N):

Standard air gas for zero calibration and O2 span calibration

 

Standard analyte/remaining N2 gas for low range span calibration and O2 zero calibration

Form verification:

Form for all components already approved

Ambient temperature:

-5 to +40° C

 

Power supply:

100± 10V AC, 50 or 60Hz

Power consumption:

Approx. 1,200 VA (excluding consent and heating tube)

Mass (weight):

Approx. 300 kg (excluding standard gas)

Finish color:

Munsell 5Y 7/1 semigloss

– 45 –

Page 53
Image 53
Fuji Labs ZSP Sample gas extracting rate Approx N/min Gas extractor, Sample inlet tube, Standard gas 3.4 N, Power supply