HP 24A, 83620A, 22A Ext Det Cal, Function Group Menu Map Description Programming Codes See Also

Models: 83620A

1 508
Download 508 pages 6.17 Kb
Page 202
Image 202
Ext Det Cal

Ext Det Cal

Function Group

Menu Map

Description

Programming Codes

See Also

ENTRY

NONE

This softkey lets you turn off (blank) the active entry area and disable the ARROW keys, rotary knob, and entry keys. When any function key (hard or soft) is pressed, the active entry area is reactivated. The yellow LED, ENTRY ON, next to ENTRY ON/OFF indicates whether the entry area is active (LED on=active).

SCPI: No specific code activates (ENTRY ON/OFF)

Analyzer:EF=off

Arrow Keys

“Entry Area” in Chapter 1.

Ext Det Cal

Function Group USER CAL

Menu Map 9

Description This softkey enables the synthesizer to act as a controller to an HP 437B power meter. This softkey causes an immediate execute on the interface bus and generates an HP-IB error if no power meter is present on the interface bus or if the synthesizer is unable to address the power meter. Use external detector calibration to characterize and compensate for an external negative diode detector used in an external leveling configuration.

Programming Codes SCPI:

CALibration:PMETer:DETector:INITiate? DIODe

CALibration:PMETer:DETector:NEXT? <num>[lvl suffix]

Analyzer:NONE

See Also

“Optimizing Synthesizer Performance” in Chapter 1.

HP 8360

Operating and Programming Reference E-5

User’s Handbook

 

Page 202
Image 202
HP 24A, 22A Ext Det Cal, Function Group Menu Map Description Programming Codes See Also, Function Group USER CAL Menu Map