(SWEEP]

Function Group SWEEP

Menu Map 7

Description This softkey lets you set a sweep time for frequency sweeps or power sweeps. The sweep time range is 10 ms to 2OOs, but the fastest sweep time is constrained by the frequency span. The fastest possible sweep can be determined automatically:

1.Press SWEEP (MENU), this reveals the sweep menu keys.

2.Select more If3 , to scroll to the next page of the sweep menu.

3.Select SwpTime Auto, to set the sweep time to automatic. The synthesizer calculates the fastest possible calibrated sweep time for any sweep span.

Whenever you press (SWEEPTIME), the active entry area displays the current sweep time and whether the sweep time is coupled to the frequency span (far right hand side displays: AUTO). If the word AUTO is not displayed then the sweep time auto function is off.

Programming Codes SCPI: SWEep[:FREQ uency]:TIME <num>[time suffix] or MAXimum] MINimum

Analyzer: ST <num> [sclms]

See Also Power Sweep

“Power Level and Sweep Time Operation,” in Chapter 1.

“Programming Typical Measurements,” in Chapter 1.

SwpTime Auto

Function Group SWEEP

Menu Map 7

S-78 Operating and Programming Reference

HP 8360

User’s

Handbook

Page 374
Image 374
HP 22A, 83620A, 24A manual Sweep, SwpTime Auto