HP 24A, 83620A, 22A manual HP User’s Handbook, Scpi Command Summary, CW or LIST Rotates around 0 Hz

Models: 83620A

1 508
Download 508 pages 6.17 Kb
Page 343
Image 343
CW or LIST Rotates around 0 Hz.

HP 8360

User’s Handbook

SCPI COMMAND SUMMARY

FREQ:MODE Affect on Slope

CW or LIST Rotates around 0 Hz.

SWEep or Rotates around the start frequency.

STEP

The *RST value is 0.

POWer:SLOPe:STATe ONOFFlO

POWer:SLOPe:STATe?

Sets and queries the power slope state. *RST value is 0.

POWer:SPAN <num>[DB] lMAXimumlMINimumlUPlDOWN

POWer : SPAN? [MAXimum Operating and Programming Reference S-45I MINimum]

The coupling equations for power sweep are exactly analogous to those for frequency sweep. Power sweep is allowed to be negative, unlike frequency sweeps. *RST value is 0.

P0Wer:STAR-t <num>~lvlsuffixllMAXimumlMINimumIUPlDOWN

POWer:STARt? [MAXimumlMINimuml

Default units and units for query response are determined by the command UNIT : POWer. The coupling equations for power sweep are exactly analogous to those for frequency sweep. Power sweep is allowed to be negative, unlike frequency sweeps.

*RST value is 0 dBm.

POWer:STATe ONOFFlO

POWer:STATe?

Sets and queries the output power on/off state. *RST value is OFF.

POWer:STEP:AUTO ONOFFlO

POWer:STEP:AUTO?

Sets and queries the function switch that controls how the power step size (Power: STEP: INCRement) is determined. if in the automatic state, then the step size is 1 dB. The *RST setting is ON.

POWer : STEP [ : INCRementI <nu.m> [DB] Manual backgroundI MAXimum Manual background1 MINimum

POWer : STEP [ : INCRementI ? [MAXimumIMINimumI

Sets and queries the power step size to be used for any node in the power subsystem that allows UP and DOWN as parameters. Setting this value explicitly causes POWer : STEP : AUTO OFF.

The *RST setting is 10 dB.

POWer:STOP Qn.un>[lvlsuffix] lMAXimumlMINimumlUPlDOWN

POWer : STOP? [MAXimumIMINimumI

Sets and queries the ending power for a power sweep. Default units and units for query response are determined by the command UNIT:POWer. The coupling equations for power sweep are exactly

Operating and Programming Reference S-45

Page 343
Image 343
HP 24A HP User’s Handbook, Scpi Command Summary, CW or LIST Rotates around 0 Hz, Operating and Programming Reference S-45