Doubler Amp Mode On

Function Group

Menu Map

Description

P O W E R

5

This softkey is applicable to instrument models with a doubler installed. The doubler has an integral amplifier whose operation is controlled by the instrument firmware. This softkey turns off the automatic mode of operation and turns on the amplifier so that it is always used. This is an unspecified mode of operation since it can cause increased harmonics and degraded dynamic range at some frequencies. This softkey has no effect on instruments without a doubler.

An asterisk next to the key label indicates that this feature is active.

Programming Codes

See Also

SCPI:

POWer:AMPLifier:STATE ON]OFF]O]l POWer:AMPLifier:STATE?

Analyzer: NONE

Dblr f&mp Menu

Dwell Coupled

Function Group FREQUENCY

Menu Map 2

Description This softkey lets you couple the dwell time for points in the stepped frequency sweep mode to the ramp sweep mode sweep time. The equation to determine the dwell time in the dwell coupled mode is as follows:

Coupled Dwell Time = (sweep time) + (number of step points)

An asterisk next to the key label indicates that this feature is active.

Programming Codes

See Also

SCPI: SWEep[:FREQ uency]:DWELl:AUTO ON]1

Analyzer: NONE

Step Sup Menu