HP 83620A, 24A, 22A manual l-42 Getting Started Advanced

Models: 83620A

1 508
Download 508 pages 6.17 Kb
Page 72
Image 72
l-42 Getting Started Advanced

using (address 13 is assumed). Refer to the menu map 8, System, for the key sequence necessary to reach softkey Meter Adrs .

Enable User Flatness Correction

13.When the operation is complete, (a message is displayed) the flatness correction array is ready to be applied to your setup.

14.To save the synthesizer parameters including the correction table in an internal register, press ISAVE) 0.

(n = number 1 through 8).

15.Disconnect the power meter/sensor and press (FLTNESS ON/OFF) [amber LED on). The power produced at the point where the power meter/sensor was disconnected is now calibrated at the frequencies and power level specified above.

l-42 Getting Started Advanced

Page 72
Image 72
HP 83620A, 24A, 22A manual l-42 Getting Started Advanced