Function Group

Menu Map

Description

INSTRUMENT STATE

NONE

This hardkey (green) causes the synthesizer to perform a short version of self-test, and initializes the synthesizer to a standard starting configuration. Two states can be defined for the standard configuration: Factory or User.

Press (PRESET) at any time to test the synthesizer and restore to a standard configuration. If the red LED adjacent to THE [PRESET) KEY (labeled INSTR CHECK) st ays on after preset, the synthesizer failed self-test; refer to “Troubleshooting Manual”.

Cycling power with the POWER switch does not have the same effect as presetting the synthesizer. Cycling power causes the synthesizer to display the programming language, the HP-IB address, and the firmware revision date. After the synthesizer displays this data, it restores its configuration to the state before power was turned off.

Programming Codes

See Also

SCPI: SYSTem:PRESet[:EXECute]

Analyzer: IP

Preset Mode Factory, Preset Mode User “Changing the Preset Parameters,” in Chapter 1. “Programming Typical Measurements,” in Chapter 1.

P-8 Operating and Programming Reference

HP 8360

User’s

Handbook

Page 282
Image 282
HP 83620A, 24A, 22A manual Scpi SYSTemPRESetEXECute