•DIAGnostics:TEST:CONTinue

<manual entry point> a string response that identifies the paragraph number in the HP 8360 Assembly-Level Repair Manual to begin the troubleshooting procedure.

DIAGnostics:TEST:CONTinue

Causes the selftest execution to continue when paused for raw data examination. Does nothing in other conditions.

DIAGnostics:TEST:DATA:DESC?

Returns the description string of the raw data examined during a selftest. It returns “ ” in other conditions.

DIAGnostics:TEST:DATA:MAXimum?

Returns the upper test limit for the raw data point examined. Returns “0” in other conditions.

DIAGnostics:TEST:DATA:MINimum?

Returns the lower test limit for the raw data point examined. Returns “0” in other conditions.

DIAGnostics:TEST:DATA:VALue?

Returns the raw data value for the raw data point examined. Returns a “0” in other conditions.

DIAGnostics:TEST:DISable (<num>)l*?IALL

Prevents the listed selftests from being selected. If ALL is sent then all of the selftests are disabled. *RST causes DIAG:TEST:ENAB ALL to execute.

DIAGnostics:TEST:ENABle CCnum>3l*?lALL

Enables the listed selftests to execute. If ALL is sent then all of the selftests are enabled. *RST causes DIAG:TEST:ENAB ALL to execute.

DIAGnostics:TEST[:EXECute] <num>

The specified selftest is executed. Normal instrument operation is suspended and the instrument state is restored upon exiting the selftest mode.

DIAGnostics:TEST:LOG:SOURce ALLIFAIL

DIAGnostics:TEST:LOG:SOURce?

Sets and queries the source for the raw data logging. ALL specifies that all raw data points are displayed. FAIL selects only those data points out of the test limits. Both commands are executable in selftest mode. After *RST, the setting is FAIL.

HP 8360

Operating and Programming Reference S-33

User’s Handbook

 