SYNTHESIZER

 

HP 4378

SYNTHESIZER

POUER flETER

 

HP ‘l37B

SYNTHFSIZER

POULR NFTFR

HICROURVE RNPLIFIER

Figure l-20.

Creating Arbitrarily Spaced Frequency-Correction Pairs in a Swept mm-wave

Environment

For this example, refer to menu map 5, POWER.

1.The equipment setup shown in Figure l-20 assumes that you have followed the steps necessary to correctly level the configuration. If you have questions about external leveling refer to earlier paragraphs titled, “Externally Leveling the Synthesizer.”

Setup Power Meter

2.Zero and calibrate the power meter/sensor.

3.Connect the power sensor to test port.

4.Enter and store in the power meter, the power sensor’s cal factors for correction frequencies to be used.

l-40 Getting Started Advanced