Internal Pulse Generator Period

Internal Pulse Mode Gate

Turns on the internal pulse mode during the positive cycle of the externally generated pulse.

Internal Pulse Mode Trigger

Triggers on the leading edge of the external pulse input.

Programming Codes SCPI: NONE, see the individual softkeys listed.

Analyzer: NONE

See Also 0,MOD a 1 so see “Modulation” and “Pulse”.

Internal Pulse Generator Period

Function Group INIOD)

Menu Map 4

Description This softkey (Option 002 only) lets you set a value for the internal pulse generator’s pulse period. The pulse is adjustable from

300 ns to 400 ms with ‘25 ns resolution. The factory preset default is 2 ms pulse period. When this feature is active, its current value is displayed in the active entry area.

Since period and rate are inversely related, if both are given values, only the last one will be applied which will cause the first one to be recalculated. Use the one that is convenient for your application. For example, if you set the pulse period, do not change the pulse rate (the synthesizer automatically adjusts the rate to match the period.)

Programming Codes SCPI: PULS:INTernal:PERiod <num>[time suffix]lMAXimum(MINimum

Analyzer: NONE

See Also 0,MOD a 1 so see “Pulse” and “Modulation”.

HP 8360

Operating and Programming Reference t-9

User’s Handbook

 

Page 234
Image 234