SECTION 7—RECLINER

2.Align threads of threaded sector with exposed threads of locking rod through cutout in recliner tube.

CAUTION

DO NOT route the cable assembly to the outside of the locking rod/recliner tube assembly as it may rub against the tires.

3.Route cable assembly to the inside of the locking rod/recliner tube assembly.

4.Position U‐bolt in middle of support plate.

WARNING

A 90° angle must be maintained to provide full contact of the threads between the sector and the locking rod (Detail “A”). Otherwise, injury to the user and/or assis- tant or damage to the equipment may result.

5.When connecting the lower cable assembly to the recliner tube, the housing must be at 90° to the frame step tube.

CAUTION

DO NOT overtighten the locknuts on the U-bolt. Otherwise, damage to the equipment may result.

6.Secure support housing/cable assembly and support plate to recliner tube with existing U‐bolt and two (2) locknuts. Torque to 4 ft‐lbs. (48 inch‐pounds).

9000 Series

62

Part No. 1076155

Page 62
Image 62
Invacare 9000 Series, 9000 XT, 9000XT Recliner, 9000 XDT, 9000 SL service manual