SECTION 7—RECLINER

Replacing Threaded Sector

NOTE: For this procedure, refer to FIGURE 7.5.

1.Disconnect the lower cable assembly. Refer to Disconnecting/Connecting Lower Cable Assembly on page 61.

2.Remove the release handle. Refer to Replacing Release Handle on page 58.

3.Remove the existing threaded sector by performing the following:

A.Thread the barrel end of the cable through the larger hole on the threaded sector.

B.Remove existing threaded sector from housing.

4.Install the new threaded sector:

A.Thread barrel end of cable assembly through the larger hole in the new threaded sector.

B.Position the barrel end of cable in smaller hole of new threaded sector.

5.Reinstall the release handle. Refer to Replacing Release Handle on page 58.

6.Squeeze the release handle and verify that the cable is activating the threaded sector.

7.If release handle does not activate threaded sector, perform the following:

A.Make sure barrel end of cable is positioned in smaller hole of new threaded sector.

B.Ensure release handle is installed properly, repeat STEP 7.

C.If release handle still does not activate threaded sector, contact Invacare technical support at the number listed on the back cover.

8.Connect the lower cable assembly. Refer to Disconnecting/Connecting Lower Cable Assembly on page 61.

Larger Hole

Barrel End of Cable

(STEPS 3,4)

(STEPS 3,4)

Threads

 

Support Housing

Threaded Sector

(STEPS 3,4)

(STEP 3)

 

 

Smaller Hole

 

(STEP 4)

Cable Assembly

FIGURE 7.5 Replacing Threaded Sector

9000 Series

64

Part No. 1076155

Page 64
Image 64
Invacare 9000 SL, 9000 XT, 9000XT Recliner, 9000 Series, 9000 XDT service manual Replacing Threaded Sector