Telephone line

Specifications

Telephone line

A telephone line (with a RJ11C modular phone jack) must be installed within

3 m (10 ft.) of the printer to support the remote diagnostics. We recommend a high-grade analog service line.

Densitometer

An X-RITE DTP— 36 Densitometer is required to calibrate the printer and for process control. It is available through Kodak. See “Accessories” on page A-1.

SCSI Cable

Asingle-ended or differential-ended SCSI cable is required to connect the host computer to the printer. Install a single-ended SCSI cable if the distance between the host computer and the printer is less than 6 meters. Use a differential-ended SCSI cable if the distance between the host computer and the printer is greater than 6 meters but less than 30 meters.

IMPORTANT: The printer is configured in single-ended mode at the factory. Before installing a differential-ended SCSI cable, you must contact your Kodak service representative to have your printer reconfigured to the differential-ended mode.

Operating Environment

NOTE: The operating environment should be free of smoke and excessive dust.

Air conditioning requirements:

The Printer generates 7,509 BTU/hr.

B-8

September 2000

Page 104
Image 104
Kodak 20R Telephone line, Densitometer, SCSI Cable, Operating Environment, Specifications, Air conditioning requirements