Specifications
Power Outlets
A separate, dedicated power line with a 208 V, 15 amp dedicated receptacle is required for the printer.
The outlets should meet the following standards:
Item | Printer Standard |
|
|
|
|
amperage | 15 amp |
|
|
wire size | 14 gauge or larger |
|
|
phase | 1 (single) |
|
|
receptacle | NEMA5— N6/15 |
|
|
impedance between ground and | less than 2 ohms |
neutral wires |
|
|
|
voltage drop across the circuit | less than 0.1 volts |
breaker |
|
|
|
distance between receptacle and | 1.5 m (5.0 ft.) |
equipment | or less |
|
|
Line Frequency
In most cases, the electric utility company is responsible for maintaining the proper frequency of the alternating current in its distribution network. If the building is supplied with power by its own