Use the calibration target provided with your scanner.
1.Click Calibrate. The Calibrate dialog box will be displayed.
2.Select Image Chain or UDDS from the
•If you select Image Chain, use the calibration target provided with your scanner. Be sure to use a good, clean calibration target. Place the calibration target in the input elevator face up and click OK.
IMPORTANT:If the White Background Accessory is installed, it must be replaced with the black background strips, and the scanner must be restarted prior to calibration.
•If you select UDDS, place a 20 lb. or 75 g/m2 bond sheet of paper in the input elevator. Do not use the calibration target and click OK.
3.Click Start. While calibration is being performed a status will be provided in the Status box.
4.Click Done when the calibration is successful.
|