UV Protection Screen: Assembly

Tools used

ρρ Allen key

Intended use

The UV protection screen in front of the speci- men plane prevents the user from looking directly into the UV rays.

Safety Notes

UV radiation can damage the eyes. Therefore, it is mandatory to install the

UV filter and adjust it correctly.

Always position the UV protection screen so that the operator can never

look directly at the light spot.

Use with one objective

1.Adjust the UV protection screen laterally using the arm.

2.Unscrew the hexagon-head screw.

3.Adjust the UV protection screen using the arm.

4.Tighten the hexagon-head screw.

5.Fasten the UV protection screen using a hexagon-head screw to the left or right side of the microscope carrier.