FluoCombi III™ – Centering the Objectives

The following configuration only has to be carried out once. It ensures that the specimen remains in the optical center when

the objective is changed.

To attain optimum results for the follow- ing settings, insert a graticule into the eyepiece. This makes locating the center vastly

easier.

Centering the objective

1.Rotate the 5× planachromat under the beam path.

2.Focus on the midpoint of the specimen.

3.Rotate the stereo planachromatic objective under the beam path.

4.Center the objective over the specimen by loosening or tightening the screw on one side of the objective holder while turning the screw on the other side in the opposite direction.

Leica M-series

User Manual

36

 

 

 