Leica IsoPro™ Motorized XY Stage

Compatibility

Leica incident light base and TL bases (TL BFDF, TL RC™, TL RCI™)

 

 

Stage size (L × W × H)

335.5 mm × 370 mm × 41.5 mm

 

 

Travel

152 mm × 102 mm (6" × 4")

 

 

Speed X,Y

30 mm/s

 

 

Motor resolution X,Y

0.25 µm

 

 

Repeatability X, Y

± 2 µm

 

 

Positioning accuracy X,Y

±20 µm over the entire positioning range

 

 

Hysteresis X, Y

30 µm

 

 

Motor

Direct current (DC) motor with encoder

 

 

Backlash

Slack-free

 

 

Adapter

All adapters for TL bases

 

 

Acceleration

S-shaped acceleration

 

 

Computer connection

USB

 

 

Control options

Leica PSC™, Leica Smart Move, Leica SmartTouch™, Leica LAS via PC

 

 

Power supply

Input: 100–240 VAC 50 / 60 Hz 1.0 A

 

Output: 15 VDC 2.66 A 40W max.

 

 

Leica M-series

User Manual

148

 

 

 

Page 148
Image 148
Leica M205 FA user manual Leica IsoPro Motorized XY Stage, Compatibility, Control options