Leica M-series User Manual 34

FluoCombi III™ – Adjusting the Parfocality

Setting up the objective nosepiece
The following configuration only has to
be carried out once. It ensures that the
parfocality remains intact, and thus the speci-
men does not need to be brought into focus
again.
1. Place a specimen under the microscope.
2. Rotate the stereo planachromatic objective
under the beam path.
3. Reduce the view so that you can see the
specimen in the overview.
4. Center the specimen under the beam path.
5. Adjust the magnification to the highest
level and focus on a flat specimen.
6. Rotate the 5× planachromat under the
beam path.
7. Unscrew the clamping screw on the
objective.
8. Reduce the view so that you can see the
specimen in the overview.