Leica EL6000 – Preparations

Setting up the instrument

Set up the Leica EL6000 so that the front is readily accessible and visible.

ρρ

The ventilation slots on the sides and back

 

panel of the device must not be covered.

ρρ

Maintain a free room of at least 150 mm on

 

the rear panel of the instrument.

Instructions for lamp replacement

The lamp contains mercury! Be abso- lutely certain to follow the handling

instructions and safety notes provided with the lamp.

Maintain cleanliness when inserting the lamp. Remove any clinging packaging

materials.

The lamp used in the instrument becomes very hot during operation and

has high internal pressure when hot. Before changing the lamp, it is mandatory to allow the instrument to cool off for at least 20 minutes.

The Leica EL6000 compact light source comes with a factory-supplied lamp,

but the lamp is not installed when shipped. This minimizes the danger of damage to the lamp during transport.

Before transporting the Leica EL6000 compact light source, you first have to

remove the lamp.

Inserting the lamp

1.Pull the power plug out of the socket, so that the EL6000 is disconnected from the power system.

2.Unscrew both screws of the cover using a 3 mm Allen key.

3. Remove the housing cover.

Leica M-series

User Manual

57

 

 

 

Page 57
Image 57
Leica M205 FA Leica EL6000 Preparations, Setting up the instrument, Instructions for lamp replacement, Inserting the lamp